Detector Block

Scintillator

BGO

Dimensions (LxWxH)

19 x 18.5 x 15 mm

Number of Crystals

7 x 7=49

In-plane crystal pitch

2.66 mm

Axial crystal pitch

2.8 mm

Detector Module

Detector dimensions (LxWxH)

155 x 38.5 x 15 mm

Module dimensions

330 x 185 x 90 mm

Number of Blocks

2 rows of 8 = 16 blocks

Photomultiplier type

Philips XP1911 (19 mm)

Number of Photomultipliers

3 rows of 9 = 27 PMT

Number of crystals

784

Whole System

Detector rings

14

Crystals per ring

448

Detector crystals

6,272

Modules

8

Photomultipliers

216

Patient opening diameter

 

 

32 cm -- brain/breast mode

36.6, 41.2, 45.8, 50.4 cm

55 cm -- whole-body mode

Detector ring diameter (octagon)

 

 

44 cm -- brain/breast mode

48.6, 53.2, 57.8, 62.4 cm

67 cm -- whole-body mode

Axial field of view

4.0 cm

Septa

No septa between rings

Transaxial rotation

45 degree span, 1 degree typical step

Energy threshold

 

323 KeV center crystal

344 KeV corner crystal

Timing window

20 ns

Data collection

3D, step and shut

Image planes

27

Gantry tilt angle

 

0 degree breast mode (patient prone)

90 degree -- whole-body/animal/brain


© 2013 The University of Texas MD Anderson Cancer Center